Biosafety Cabinet Class II /Type B2/ (product, operator and environment protection)
BAVp-01-«Laminar-S» - 1,2 (230.120)
Attention!
The connection to the exhaust system with productivity not less than 1015 m3/h is obligatory!
The laboratory should be equipped with the forced ventilation with productivity not less than 1000 m3/h.
The cabinet is designed for protection of the operator, product and environment from pathogenic agents and microorganisms of groups II-IV pathogenicity,transmitted by the air-drop way
when carrying out microbiological operations with the use of toxic chemicals.
- No air recirculation in the working chamber
- The contaminated air is effectively cleared and completely removed into the external exhaust system
- Due to the difference of air in-flow and out-flow, in the front window there is an air veil preventing the pathogenic agent ingress into the laboratory air environment
- contaminated air
- inflow
- purified air
| External dimensions, mm (w/d/h) |
1170Ñ…740Ñ…2010 |
| Working chamber dimensions,mm (w/d/h) |
1105Ñ…610Ñ…670 |
| Incoming air volume in the working chamber, m3/h |
550 |
| Air flow velocity in the working chamber
at a distance of 150 mm from filter, m/sec |
0,35 ± 20% |
| Power consumption,W |
350 |
Weight, kg |
180 |
|
|
|
|
|