Biosafety Cabinets Class II
model 230.120

     

     

Biosafety Cabinet Class II /Type B2/
(product, operator and environment protection)

     

BAVp-01-«Laminar-S» - 1,2 (230.120)

     

Attention!
The connection to the exhaust system with productivity not less than 1015 m3/h is obligatory! The laboratory should be equipped with the forced ventilation with productivity not less than 1000 m3/h.

The cabinet is designed for protection of the operator, product and environment from pathogenic agents and microorganisms of groups II-IV pathogenicity,transmitted by the air-drop way when carrying out microbiological operations with the use of toxic chemicals.

     

     

Distinctive features

  • No air recirculation in the working chamber
  • The contaminated air is effectively cleared and completely removed into the external exhaust system
  • Due to the difference of air in-flow and out-flow, in the front window there is an air veil preventing the pathogenic agent ingress into the laboratory air environment

Air Flow Scheme

Laminar air flow cabinet

     

     

     

     

- contaminated air

- inflow

- purified air

     

     

     

     

     

     

     

Technical parameters

External dimensions, mm (w/d/h) 1170Ñ…740Ñ…2010
Working chamber dimensions,mm (w/d/h) 1105Ñ…610Ñ…670
Incoming air volume in the working chamber, m3/h 550
Air flow velocity in the working chamber at a distance of 150 mm from filter, m/sec 0,35 ± 20%
Power consumption,W 350
Weight, kg 180