Saturday, 19.05.2012

Welcome!

Biosafety Cabinet Class I


model 110.120

     

     

Biosafety Cabinet Class I
(operator and environment protection)

     

BAV -"Laminar-S" - 1,2 (code 110.120)
BAV -"Laminar-S" - 1,5 (code 110.150)

     

Functioning
BIOSAFETY CABINET CLASS I is designed for operator and environment protection from materials hazardous for operator’s health.

Opearation principle
The operation is based on the forced removal of hazardous substances from the working zone through the air up-flow,with high-effective air filtration and its exhaust into the environment. Working with 2 agents, there is no risk of their cross-contamination owing to unidirectional (laminar) air up-stream.

     

     

Distinctive features

  • HEPA filters are made by a German firm whose quality management system is certified ISO 9001:2000; they were tested under EN 1882-1 Eurostandard and provide air purification with efficiency of 99,9995 % for 0,3 µm particles
  • HEPA filter is clamped with springs providing the filter tightness for the whole operation life.
  • Ø205 flange for connection to external exhaust system with the check valve at the air output point.
  • The electronic control system provides automatic control of air flow velocity regardless the filter impurity; there is indication of necessity of filter replacement.

Options

  • Platform
  • Additional Sockets
  • Gas and vacuum supply chokes

Air Flow Scheme

Laminar air flow cabinet

     

     

     

     

- contaminated air

- inflow

- purified air

     

     

     

     

     

     

     

Technical parameters

Model 110.120 110.150
External overall dimensions, mm (w/d/h) 1170x685x1205 1470x685x1205
Working chamber dimensions,mm (w/d/h) 1105x620x670 1405x620x670
Platform dimensions,mm (w/d/h) 1165x660x660 1465x660x660
Working zone illumination,Lux 1000 1000 1000 1000
UV-lamp capacity, W 30 30
Power consumption, W 510 510
Pre-filter class according to R 51251-99 G4 G4
HEPA filter class according to R 51251-99 H14 H14
Quantity of filter steps 2 2
Filter efficiency for 3 µm particles, % 99,9995 99,9995
Air flow velocity in the operator window, m/sec 0,7 ± 20% 0,7 ± 20%
Air exhaust from the working chamber, m3/h 750 950
Weight, kg 115 135