LAMINAR AIR FLOW CABINET (HOOD) with vertical down air flow (product protection)
BAVnp-01-"Laminar-S" - 1,2 (code 410.120) BAVnp-01-"Laminar-S" - 1,5 (code 410.150) BAVnp-01-"Laminar-S" - 1,8 (code 410.180)
Functioning
LAMINAR AIR FLOW CABINETS are designed for creation of abacterial dust free air space.
LAMINAR AIR FLOW CABINETS are used for work with low-risk substances and materials, when protection of working material from environment
is required or work with item requires a sterile working zone. Cabinets are used for equipment of local workplaces in medical,
pharmaceutical and other institutes with high requirements for air cleanliness in the working zone.
- HEPA filters are made by a German firm whose quality management system is certified ISO 9001:2000;they were tested under
EN 1882-1 Eurostandard and provide air purification with efficiency of 99,9995 % for 0,3 µm particles.
- HEPA filter is clamped with springs providing the filter tightness for the whole operation life.
- Filter replacement is made through two special side openings.
- The electronic control system provides automatic control of air flow velocity regardless the filter impurity;
there is indication of necessity of filter replacement.
- Electronic shield panel provides easy exploitation and disinfection
- All electric control systems are not placed in the main working chamber
- Window rising mechanism allows fixing window in two positions which makes the working chamber and the window
easy to clean from both sides with disinfectant solutions.
- The illumination block is taken out of the working chamber and does not initiate air flow turbulence.
- Coloring of the cabinet is made with the powder enamel proof for disinfectant solutions.
- Side glass provide additional view at while working
- Platform
- Extra Sockets
- Gas and vacuum supply chokes
- contaminated air
- inflow
- purified air
| Filter efficiency for 0,3 µm particles, % |
99,9995 |
| Quantity of filtration steps |
2 |
| Pre-filter class, R 51251-99 |
G4 |
| HEPA filter class, R 51251-99 |
H14 |
| Air cleanliness class in the working zone for 0,5 µm particles |
5ISO |
| Noise pressure, dB, no more |
57...60 |
| Illumination of working zone, Lux, no less |
1000 |
| UV-lamp,W |
30 |
| Model BAVnp-01-«Laminar-S» |
410.120 |
410.150 |
410.180 |
| External overall dimensions, mm (w/d/h) |
1170х685х1115 |
1470х685х1115 |
1770х685х1115 |
| Working chamber dimensions,mm (w/d/h) |
1105х620х670 |
1405х620х670 |
1705х620х670 |
| Platform dimensions, mm (w/d/h) |
1165х660х660 |
1465х660х660 |
1765х660х660 |
| Recirculation air volume, m3/h |
710 |
905 |
1100 |
| Air flow velocity in the working chamberat a distance
of 150 mm from filter, m/sec |
0,45 ± 20% |
0,45 ± 20% |
0,45 ± 20% |
| Power consumption,W |
210 |
600 |
550 |
| Weight, kg |
145 |
160 |
175 |
|
LAMINAR AIR FLOW CABINET (HOOD) with vertical air down flow (product protection) for microelectronics
BAVnp-01-”Laminar-S” - 1,2 - ME (code 450.120) BAVnp-01-”Laminar-S” - 1,2 - ME (code 250.120)
Application
Assembly lines in instrument engineering and microelectronics; PCB and high-efficiency semiconductor indicator production, for high-class clean working zones in clean room
of low-class cleanliness. Laminar air flow cabinets provide local clean working zone inside clean laboratories.
There are two types of these cabinets:
- BAVnp-01-"Laminar-S"-1,2-ME (code 450.120)
- BAVnp-01-"Laminar-S"-1,2-ME (code 250.120)
- Moving tool storage drawer
- Corrosion-proof table surface made of unique dielectric plastic
- Additional sockets for different voltages (5V, 36V, etc.)
Laminar air flow cabinet provides sterile air environment protecting product from contaminated environment.
It has a front glass cut for microscope and is used for assembly work
Laminar air flow cabinet provides sterile air environment in the working zone and protects an operator from toxic,
volatile andsmelly substances and liquids, completely removing them from the working zone. The cabinet requires external air exhaust system
Distinctive features of BAVnp-01-"Laminar-S"-1,2-ME (code 250.120)
- Complete air exhaust from working zone to the external forced air ventilation
- Flat removable table surface made of special plastics
| Model BAVnp-01-«Laminar-S»-ME |
450.120 |
250.120 |
| External overall dimensions, mm (w/d/h) |
1170х695х1210 |
1170х740х1305 |
| Working chamber dimensions,mm (w/d/h) |
1105х620х670 |
1105х610х670 |
| Recirculation air volume, m3/h |
710 |
550 |
| Air flow velocity in the working chamberat a distance
of 150 mm from filter, m/sec |
0,45 ± 20% |
0,35 ± 20% |
| Power consumption,W |
600 |
600 |
| Weight, kg |
145 |
160 |
|
LAMINAR AIR FLOW CABINET (HOOD) with vertical air down flow (product protection) Model Vis-A-Vis
BAVnp-01-"Laminar-S" - 1,2 (code 420.120) BAVnp-01-"Laminar-S" - 1,5 (code 420.150) BAVnp-01-"Laminar-S" - 1,8 (code 420.180)
Functioning
The main feature of LAMINAR AIR FLOW CABINET (HOOD) with vertical air down flow
(model Vis-A-Vis) is itspossibility of simultaneous work of two operators sitting opposite
each other.
- Stationary UV-unit for working chamber
- Flat non-removable table surface made of stainless steel
- Heat-strengthened (rising) front and side glass
- contaminated air
- inflow
- purified air
| Model BAVnp-01-«Laminar-S» |
420.120 |
420.150 |
420.180 |
| External overall dimensions, mm (w/d/h) |
1170х680х1150 |
1470х680х1150 |
1770х680х1150 |
| Working chamber dimensions,mm (w/d/h) |
1105х560х670 |
1400х560х670 |
1700х560х670 |
| Platform dimensions, mm (w/d/h) |
1165х565х660 |
1465х565х660 |
1765х565х660 |
| Recirculation air volume, m3/h |
710 |
905 |
1100 |
| Air flow velocity in the working chamberat a distance
of 150 mm from filter, m/sec |
0,45 ± 20% |
0,45 ± 20% |
0,45 ± 20% |
| Power consumption,W |
210 |
600 |
600 |
| Weight, kg |
160 |
175 |
180 |
|
|